Multifocal Electroretinography (mfERG)

Multifocal ERG is a visual electrophysiology test of the macula that can assess macular dysfunction and monitor the progression of conditions such as age-related macular degeneration (AMD), toxicity, and others. This test can help discriminate between various retinal diseases and can also be used to rule out organic vision loss. 

mfERG multifocal ERG result in software

Taking the mfERG test

The mfERG test may be performed on anyone who can reliably fixate their eyes on a target. 

During this test, the patient sits with their chin on a chinrest in front of a specialized monitor. Their face and neck should be relaxed as they focus on a fixation target in the center of the monitor. When the test begins, the multifocal stimulus starts to alternate at high frequency while retinal responses are recorded binocularly. The procedure consists of short recording periods interspersed with pauses for blinking. This test takes approximately four minutes and utilizes noninvasive corneal electrodes to provide superior signal quality. 

Multifocal ERG stimulus

The multifocal ERG stimulus is comprised of several scaled hexagons. Each hexagon changes pseudo randomly from black to white at a rapid speed. A separate signal is recorded in response to each hexagon. To select the best stimulus size for your patient, consider the tradeoffs between spatial resolution, signal to noise ratio, and examination time.

61 Hexagon Multifocal

This is the most common multifocal test for the central 50 degrees of the visual field. It is complete with normative data for interpretation. 

103 Hexagon Multifocal

This mfERG uses 103 hexagons across 50 degrees of the visual field in order to detect the smallest defects. 

19 Hexagon Multifocal

This test can be used with smaller stimulators, like Envoy, for an assessment of macular function.  

Interpret the results of the mfERG test

Multifocal ERG results are displayed in a few different graphs, each showing a different representation of retinal function.

Response Density

A 3D topographical representation that illustrates the overall signal strength from each area of the retina.  

Traces

Traces showing the response that correlates with each hexagonal stimulus element.  

Ring Averages

Averages of the mfERG traces from concentric parafoveal regions, with clearly demarcated normative values. 

Ring electroretionography (rERG) test

Toxicity screening with ring ERG (rERG)

The ring electroretinography (rERG) method is a variation of multifocal electroretinography (mfERG) designed to screen for retinal toxicity. In the rERG method, the stimulus zones are based on true annular rings rather than a 61-hexagon pattern. This shortens the test time from over 5 minutes to around 1 minute, significantly reducing the test burden.

Screen for retinal toxicity from

  • Hydroxychloroquine
  • Chloroquine
  • Tamoxifen
  • Vigabatrin

The rERG stimulus pattern has four rings across the central 50° of vision. The ratios of the amplitude densities are calculated from the second or third ring, relative to the fourth (outermost) ring. Reference normative data indicate that measurements using the second ring are suitable for testing subjects of European, African, and Hispanic descent due to their typical parafoveal pattern of toxicity. Measurements using the third ring are suitable for testing subjects of Asian descent, due to their typical peripheral presentation of retinal toxicity damage.