The E3 console is the heart of the Espion Electrophysiology system. The E3 console features a fanless, hard-anodized aluminum chassis, real-time DSP, USB control, an isolated power supply and 5 channel isolated differential amplifiers. The E3 connects to any Windows PC and can be wirelessly connected over standard wi-fi Ethernet. The E3 includes easy-to-use software to run standard ISCEV tests as well as setup research experiments in an intuitive manner.
Espion E3 Detailed Specifications
Espion E3 Electroretinography System
The Espion E3 is a full featured, modular electrophysiology
system capable of generating visual stimuli for any stimulator.
Together with Diagnosys’ industry-leading stimulators, it is
capable of performing not only all ISCEV standard ERG tests,
but any visual electrophysiology tests, even those
previously possible only with custom lab-built equipment.
Design Features
Sleek, modern design
Hardened, machined aluminum case
Multiple PC configurations including: Touchscreen,
Laptop, Rugged Laptop, Tablet and Desktop PC
Features
Advanced real-time DSP based acquisition
system with integrated power isolation
5 integral computer-controlled fully-isolated differential
amplifiers with industry-first 32-bit ADCs, all digital low,
high and band pass filters, DC input, (not AC coupled)
and new, light-guided impedance measurement system
2 TTL compatible isolated trigger input and output BNCs
Power on and status indicator LEDs
USB communication
Direct pattern generator connection
Isolated auxiliary line outlets
Espion Software
Includes ISCEV standard test protocols and the ability to
create new protocols for custom testing
Specialized animal testing protocols available
Industry standard SQL database using an advanced
interface for comprehensive data storage and retrieval
SQL based Patient Management System
Microsoft Office™ integration for intuitive, drag and drop report generation
Data Acquisition
The data acquisition system is based around a fast DSP microprocessor that
controls and synchronizes data acquisition and stimulator timing independently
from the PC. The system is comprised of 5 differential DC coupled amplifiers (pre-
vent AC amplifier lockup) with the following specification:
Industry-first 32 bit resolution (<1nV LSB)
Fully isolated
Input range +/- 2.5V max; +/-625 mV default
Automated internal calibration
Integral anti-alias filter
Differential channel input gain matching better than 0.01%
Ultra low noise <1uV RMS
Input impedance greater than 109 ohms
CMRR greater than 110dB at 50/60Hz
4kHz internal sample frequency
Automatic impedance measurement measures each input including ground,
not just each channel
All filtering is performed digitally
Control Computer Option
Touchscreen panel PC, Rugged Laptop, Laptop, or Desktop PC configured to
current highest hardware standard available at time of purchase
USB optical mouse and keyboard included
Power Isolation
The Espion Profile System incorporates transformers required to isolate ancillary
equipment such as a monitor or printer so they will meet medical leakage current
standards. No cumbersome external transformers are needed.
Unique lead testing system uses illuminated receptacles to guide lead insertion and indicate electrode contact quality.